Applied Materials 0010-21246 Part, Perkin Elmer Lambda 35, side by side. Every value shown is drawn from the cited encyclopedia entries.

Side-by-side comparison of selected equipment models
Attribute
0010-21246 PartApplied Materials
Lambda 35Perkin Elmer
OEMApplied MaterialsPerkin Elmer
Category——
Wafer size——
Process node——
Introduced——
Production run——
Lifecycle——
Lifecycle milestones——
Control system——
Generation——
Family—Perkin Elmer Lambda 35
Cited variants—
Specifications
Instrument type—Scanning double-beam spectrometer for the UV/Vis range; operation by PC[1]
Wavelength range—190 nm to 1100 nm[1]
Spectral bandwidth—0.5 nm, 1 nm, 2 nm, 4 nm (selectable fixed slit widths)[1]
Detector—Photodiodes, one for the sample beam and one for the reference beam[1]
Radiation sources—Prealigned deuterium and halogen lamps[1]
Photometric range—Transmission 0% to 100%; absorbance -6.000 to 6.000; concentration units 1 to 9999[1]
Photometric accuracy—Absorbance ±0.003 at Absorbance = 1 at 440 nm, 546.1 nm and 635 nm; absorbance ±0.015 at Absorbance = 1 at 257 nm and 350 nm[1]
Stray radiation—Transmission <0.02% at 220 nm, 340 nm and 370 nm; absorbance >2 at 200 nm with potassium chloride solution against distilled water[1]
Baseline stability—Absorbance <0.0003 per hour at 500 nm after warmup[1]
Beam center height—15 mm above cell holder bottom[1]
Optical pathlength in sample compartment—121 mm[1]
Interface—One RS 232 C interface (serial) for connecting a PC[1]
Use cases—Measurements on liquids, solids, pastes and powder samples[2]
Modes of operation—Scanning, wavelength program, time-drive, rate, quant, scanning quant[3]
Type—Scanning double-beam spectrometer for the UV/Vis range; operation by PC[1]
Dimensions—Width: 650 nm; Height: 260 nm; Depth: 560 nm[1]
Mass—26 kg approx.[1]
Power requirements—100 V to 240 V AC, 50/60 Hz; 250 VA[1]
Ambient operating temperature—15 ºC to 35 ºC[1]
Humidity range—20% to 80% relative humidity without condensation[1]
Wavelength accuracy—±0.1 nm[1]
Wavelength reproducibility—±0.1 nm[1]
Lamp change—Automatically at 326 nm (software selectable over the whole wavelength range)[1]
Scan speeds—7.5, 15, 30, 60, 120, 240, 480, 960, 1920 and 2880 nm/min[1]
Baseline linearity—2 nm slit: Absorbance ±0.0005 (corrected: 200 nm to 1100 nm, scan speed 240 nm/min, smooth 2)[1]
Baseline noise—2 nm slit: Absorbance < 0.00006 RMS; Absorbance < 0.0002 peak-to-peak 3 min at Absorbance = 0, wavelength 500 nm and response 2 s[1]
Data output—One RS 232 C interface (serial), for connecting a PC[1]
Optical system—All-reflecting optical system with silica-coated optical components and a holographic concave grating with 1053 lines/mm in the center[1]
Beam cross-section—0.5 nm slit ca. 0.25 mm x 7 mm; 1 nm slit ca. 0.5 mm x 7.5 mm; 2 nm slit ca. 1 mm x 7.5 mm; 4 nm slit ca. 2 mm x 7.5 mm[1]
Grating—Holographic concave grating with 1053 lines/mm in the center[1]
Baseline stability (drift)—Absorbance < 0.0003 per hour at 500 nm after warmup[1]