Waferpedia

Edwards

IH 1000

Lab & Support EquipmentEdwards IH 1000 family
Research Quality: 50% complete

The Edwards iH1000 is a dry vacuum pumping system that uses an HCDP80 dry pump with an HCMB1000 mechanical booster pump, achieving a peak pumping speed of 1000 m³/h at 60 Hz. The iH1000 is part of the iH series of dry pumping systems from Edwards (formerly BOC Edwards).[1]

IH 1000 — Inventory photo
Fig. 01IH 1000Inventory photo[2]

Power

1000 m³/h[1]

Vacuum

0.75 litres (typical)[1]

How it works

The iH dry pumping system uses a HCDP dry pump as the primary pumping stage. For the iH1000, this is supplemented by a HCMB mechanical booster pump to increase pumping speed at lower pressures.[1]

The system is designed with a temperature control system for both the HCDP and HCMB pumps, and includes connections for nitrogen purge and cooling water to manage process gases and thermal load. The pumping system is controlled by an electronics module and a pump display terminal.[1]

Dry pumps in this class use claw or screw mechanisms to compress gas without lubricants, making them suitable for clean vacuum environments. The iH series includes features such as gas modules, exhaust systems, and safety sensors.[1]

What do the numbers mean?

Power & electrical10

Typical peak pumping speed (60 Hz)
1000 m³/h[1]
Accurate?
Typical peak pumping speed (50 Hz)
950 m³/h[1]
Accurate?
Typical ultimate vacuum (60 Hz)
1×10⁻³ mbar[1]
Accurate?
Typical ultimate vacuum (50 Hz)
2×10⁻³ mbar[1]
Accurate?
Full load current (460 V, 60 Hz)
13 A[1]
Accurate?
HCMB motor rating
2.6 kW (60 Hz)[1]
Accurate?
Maximum power input to both pump motors (60 Hz)
7.8 kW[1]
Accurate?
Typical ultimate vacuum with shaft-seals purge only (60 Hz)
1 x 10^-3 mbar[1]
Accurate?
Typical ultimate vacuum with shaft-seals purge only (50 Hz)
2 x 10^-3 mbar[1]
Accurate?
Electrical supply voltage options
200-208/380-415 V at 50 Hz; 200-208/230/460 V at 60 Hz[1]
Accurate?

Vacuum & pumping5

HCDP pump oil charge (gearbox)
0.75 litres (typical)[1]
Accurate?
HCMB pump oil charge (gearbox)
0.78 litres (typical)[1]
Accurate?
Maximum cooling-water supply pressure
100 psig (6.9 bar)[1]
Accurate?
Nitrogen supply pressure range
20 to 100 psig (2.4 to 7.9 bar)[1]
Accurate?
Pump configuration
HCDP80 dry pump with HCMB1000 mechanical booster pump[1]
Accurate?

Dimensions & facilities1

Mass
430 kg[1]
Accurate?

Configuration & options9

Without Invrt[4]
Accurate?
Fully refurbished[4]
Accurate?
With Invrt[2]
Accurate?
Rear[3]
Accurate?
Noise level (at 1 m with enclosure)
<69 dB(A)[1]
Accurate?
Inlet connection
ISO100 bolted[1]
Accurate?
Outlet connection
NW40[1]
Accurate?
Cooling-water minimum flow rate
4 l·min⁻¹[1]
Accurate?
Noise level (measured at 1 m with enclosure panels fitted)
<69 dB(A)[1]
Accurate?

Vintage & configurations

Documented models & variants

DesignationGenerationVintageChangesSource
IH 1000 Pump - Without Invrt - Fully refurbished——Described as without inverter and fully refurbished.Equipment inventory listing[4]
IH 1000 Pump - With Invrt——Described as with inverter.Equipment inventory listing[2]
IH 1000 Pump - Rear——Described as rear.Equipment inventory listing[3]
iH1000F——Includes Breach Sensor for leak detection during run 'til crash.nanofab.utah.edu[1]
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Typical peak pumping speed (60 Hz)1000 m³/h[1]
  • Typical peak pumping speed (50 Hz)950 m³/h[1]
  • Typical ultimate vacuum (60 Hz)1×10⁻³ mbar[1]
  • Typical ultimate vacuum (50 Hz)2×10⁻³ mbar[1]
  • HCDP pump oil charge (gearbox)0.75 litres (typical)[1]
  • HCMB pump oil charge (gearbox)0.78 litres (typical)[1]
  • Full load current (460 V, 60 Hz)13 A[1]
  • HCMB motor rating2.6 kW (60 Hz)[1]
  • Maximum power input to both pump motors (60 Hz)7.8 kW[1]
  • Maximum cooling-water supply pressure100 psig (6.9 bar)[1]
  • Nitrogen supply pressure range20 to 100 psig (2.4 to 7.9 bar)[1]
  • Typical ultimate vacuum with shaft-seals purge only (60 Hz)1 x 10^-3 mbar[1]
  • Typical ultimate vacuum with shaft-seals purge only (50 Hz)2 x 10^-3 mbar[1]
  • Electrical supply voltage options200-208/380-415 V at 50 Hz; 200-208/230/460 V at 60 Hz[1]
  • Pump configurationHCDP80 dry pump with HCMB1000 mechanical booster pump[1]

Where are the manuals?

Publicly hosted documents referencing this tool, linked at their original location. Hosted by the linked institutions — availability may change.

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What pump stages does the IH 1000 use?

The IH 1000 combines an HCDP dry pump with an HCMB mechanical booster pump. The HCMB booster is fitted on the iH600 and iH1000 systems only.[1]

What utility services does the IH 1000 require?

The IH 1000 requires a nitrogen supply, cooling-water connections, and an electrical supply. Optional accessories include an enclosure extraction fan and alternative nitrogen and cooling-water connection kits.[1]

Does the IH 1000 comply with the ATEX directive?

The iH dry pumping systems are declared compliant with the ATEX directive as Category 3GD equipment for internal atmospheres only.[1]

What maintenance does the IH 1000 require?

Maintenance includes checking and changing the oil in the HCDP pump and the HCMB pump on iH600 and iH1000 systems. Maintenance also includes checking purge-gas flow rates, inspecting and cleaning the exhaust pipe, and zeroing the gas-module flow transducer.[1]

Not publicly documented

The following facts about the IH 1000 are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the IH 1000 are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • The control-system platform and OS era of the IH 1000 are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the IH 1000 are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the IH 1000 is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the IH 1000 are on record.

    Answerable by: an OEM parts catalog or a service engineer

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (4)Every fact above is drawn from these public sources
  1. [1]nanofab.utah.edu — nanofab.utah.edunanofab.utah.edu
  2. [2]Inventory photo
  3. [3]Inventory photo
  4. [4]inventory listing
Was this page accurate?

Last updated Oct 4, 2026.

Compare with similar tools

View all →