Waferpedia

Comparison ledger

0010-03254 E ChuckversusTSE 12 A Thermal Shock

Applied Materials 0010-03254 E Chuck, ESPEC TSE 12 A Thermal Shock, side by side. Every value shown is drawn from the cited encyclopedia entries.

Side-by-side comparison of selected equipment models
Attribute
0010-03254 E ChuckApplied Materials
OEMApplied MaterialsESPEC
Category——
Wafer size——
Process node——
Introduced——
Production run——
Lifecycle——
Lifecycle milestones——
Control system——
Generation——
Family—ESPEC TSE 12 A Thermal Shock
Specifications
Phase—3 Phase[1]
Frequency—60 Hz[1]
Interior Volume—0.4 cu. ft. (11 Liters)[2]
Interior Dimensions (W x D x H)—12.5" x 9.1" x 5.8" (320 x 230 x 150 mm)[2]
Exterior Dimensions (W x D x H)—26" x 42" x 79" (NEC compliant version) (680 x 1050 x 2007 mm)[2]
Temperature Range—-65 to 200°C (-85 to 392°F)[2]
Power—208V (±5%) 3ph 60Hz 30 Service Amps (26 FLA)[2]
Maximum Sound Level (1 meter away)—60 dBA[2]
Temperature Fluctuation (at control sensor)—±0.5°C[2]
Refrigeration System—Cascade 2hp Scroll[2]
Transfer Mechanism—Vertical-lift, direct-drive system for reliable transfer between zones[3]
Maximum Load—17 lb. (4 kg)[3]
Specimen Baskets—Two specimen baskets[3]
Cable Port—Tube for cables to test load, 2" diameter[3]
Specimen Power Supply Terminal—Included[3]
Refrigeration Type—Non-CFC cascade refrigeration system using 2-hp compressors[3]
Air Circulation—Two sirocco blower motors circulate the air (one per zone)[3]
Controller—ESPEC's Exclusive P-310 Programmer with 7" color touch-screen[3]
Programs—20 user-definable programs can be saved with alphanumeric names[3]
Connectivity—Ethernet with browser-based access is standard, plus USB[3]
Utility Requirement—The only utility required is 208V power[3]
Width—Only 26 inches wide[3]
Compliance—Meets Mil-Std 883 1010.7/1010.8 thermal cycling requirements with 4 kg load; also can be used for IEC 61300-2-47, JEDEC JESD47E, JESD22-A104, and others[3]
Test Performance Example—Test: Mil-Std 883 1010.7 Range: -65 to 150°C Recovery: 15 min. (worst-case on-product) Load: 6 lb of IC chips[2]
Test performance example—Mil-Std 883 1010.7; range -65 to 150°C; recovery 15 min. (worst-case on-product); load 6 lb of IC chips[2]
Heat-to-room peak—17,000 BTU/Hr.[2]
Maximum sound level—60 dBA (1 meter away)[2]
Temperature fluctuation—±0.5°C at control sensor[2]
Maximum load—17 lb.[2]
Cable tube diameter—2 in.[2]
Display—7 in. color touch-screen[2]
Sources (3)Every fact above is drawn from these public sources
  1. [1]inventory listing
  2. [2]espec.comespec.com